Publication in refereed journal
香港中文大学研究人员 ( 现职)
孙德生教授 (医学院) |
麦耀光教授 |
严秉泉教授 (外科学系) |
全文
数位物件识别号 (DOI) http://dx.doi.org/10.1177/021849239700500114 |
引用次数
Scopushttp://aims.cuhk.edu.hk/converis/portal/Publication/0Scopus source URL
其它资讯
摘要We report an unusual case of severe mechanical hemolysis after implantation of a 17-mm Rashkind patent ductus arteriosus occluder in a young adult female with a large ductus. The device failed with conservative management, including implantation of a second occluder, and surgery was required to close the ductus. This occluder may not be suitable for closure of a large patent ductus arteriosus.
着者Ho J.K.S., Yim A.P.C., Lee J.F.Y., Tse L.K.K., Mak G.Y.K., Low J., Sanderson J.
期刊名称Asian Cardiovascular and Thoracic Annals
出版年份1997
月份12
日期1
卷号5
期次1
出版社SAGE Publications
出版地United States
页次56 - 58
国际标準期刊号http://aims.cuhk.edu.hk/converis/portal/Publication/0218-4923
语言英式英语